Magnetocardiography, a non-invasive technique for diagnosing heart disease, relies on precise measurements of the tiny magnetic fields produced by the heart’s electrical activity, but these measurements can be compromised by the presence of metallic implants. Ho-Seong Lee from Korea University, alongside Jae-Hyun Ahn and Yong-Hwan Kim from AMCG, investigated how common implant materials influence the accuracy of magnetocardiography readings. Their work addresses a growing concern as the number of patients with implants increases, and importantly, establishes that standard implant materials have minimal impact on diagnostic measurements, even under challenging experimental conditions. The team’s findings, achieved using a highly sensitive 96-channel magnetocardiography system, reassure clinicians and patients that this vital diagnostic tool remains reliable for a broad patient population, although further research into the effects of micromotion within the body is still needed.
Researchers investigated how common implant materials influence the accuracy of these measurements, addressing a growing concern as the number of patients with implants increases.
The team’s work establishes that standard implant materials have minimal impact on diagnostic measurements, even under challenging experimental conditions. Their findings, achieved using a highly sensitive 96-channel magnetocardiography system, reassure clinicians and patients that this vital diagnostic tool remains reliable for a broad patient population, although further research into the effects of micromotion within the body is still needed.
Implant Material Impacts Magnetocardiography Signal Quality
Scientists conducted a detailed analysis of the impact of implant materials on magnetocardiography (MCG) measurements, a non-invasive technique for diagnosing cardiac disease. The research team utilized a 96-channel MCG system, housed within a magnetically shielded room to minimize external interference, and focused on titanium-6Aluminum-4Vanadium ELI (Ti-6Al-4V ELI), a commonly used implant material.
Researchers deliberately created extreme conditions, positioning a sample as close as possible to the sensors, to maximize any potential magnetic disturbance. Measurements revealed a minimal noise increase of approximately 0.7 fT/ √Hz, remaining within established sensitivity criteria for accurate MCG measurements. This breakthrough confirms the feasibility of performing MCG examinations on patients with metallic implants without compromising diagnostic accuracy, potentially expanding access to this valuable cardiac assessment tool. Further investigation will focus on verifying the absence of magnetic fields generated by micromotion of the implant material within the human body.
Minimal Implant Interference In Magnetocardiography
This research successfully investigated the potential impact of common implant materials on magnetocardiography (MCG) measurements, a non-invasive technique for diagnosing cardiac disease. Scientists conducted experiments using a 96-channel MCG system and a magnetically shielded room, focusing on Ti6Al4V ELI titanium alloy as a representative implant material.
The team deliberately placed the material in close proximity to the sensors, simulating worst-case conditions to assess potential interference. The results demonstrate that even under these extreme conditions, the increase in noise was minimal and remained within acceptable limits for accurate MCG readings. This finding suggests that the presence of typical metallic implants poses little to no significant disruption to the diagnostic process, offering reassurance for the growing number of patients with such devices. While the study confirms minimal interference, the researchers acknowledge that the dynamic environment of a living body, specifically micromotion of the implant within tissue, requires further investigation.
👉 More information
🗞 Study of the Influence of Implant Material on Magnetocardiography Measurements Using SQUID Sensors
🧠 ArXiv: https://arxiv.org/abs/2512.20859